Ventricular arrhythmia discriminator programming and the impact on the incidence of inappropriate therapy in patients with implantable cardiac defibrillators

Srivatsa, Uma N.; Hoppe, Bobbi L.; Narayan, Sanjiv; Feld, Gregory K. and Birgersdotter-Green, Ulrika (2007) Ventricular arrhythmia discriminator programming and the impact on the incidence of inappropriate therapy in patients with implantable cardiac defibrillators. Indian Pacing and Electrophysiology Journal, 7 (2). pp. 77-84. ISSN 0972-6292

Full text available as:

PDF - Requires Adobe Acrobat Reader or other PDF viewer.
306 Kb

Abstract

Background: The incidence of inappropriate therapy from implantable cardioverter defibrillators (ICDs) has been reduced by programming ventricular arrhythmia discriminators (VAD) on at the time of implant. Objective: To determine which VAD is most effective in preventing inappropriate therapy. Methods and Results: Dual chamber ICD (n=48) or cardiac resynchronization therapy defibrillator (CRT-D) (n=55) implantation was performed in 103 patients (M=94, F=9). Patients were followed prospectively for therapy events (shock or anti-tachycardia pacing) for a mean 362±289 days. Events were correlated with clinical characteristics and VAD programming. Of the 103 pts followed, 11 received inappropriate therapy (IT), 15 received appropriate therapy (AT), and 77 received no therapy (NT). In the AT and IT groups, a total of 207 events (ATP=171, shock=36) were observed. A total of sixty-four electrograms (EGMs) were analyzed. Programming VADs "ON" versus "OFF" reduced the incidence of IT events compared to those receiving AT or NT events (p<.01), with a trend in fewer patients receiving IT (31.3% "ON" vs 55.6% "OFF", p = 0.131). Programming atrial fibrillation (AF) detection ON resulted in fewer patients receiving IT compared to those receiving AT or NT (3.6% vs 19%, p<.05). Furthermore, programming AF or AFL algorithms "ON", resulted in overall fewer episodes of IT therapy (p<.01). Conclusions: AF or AFL discriminators significantly reduced the incidence of IT, and were predominantly responsible for the benefits from VAD programming observed in this study. Activating these features as part of routine ICD or CRT-D programming may provide a simple and effective alternative to the use of more complex and multiple VAD strategies.

EPrint Type:Article
Uncontrolled Keywords:Implantable Cardiac Defibrillators, Inappropriate Therapy, Ventricular Arrhythmia Discriminators
Subjects:Natural Sciences > Mathematics
-Journal Repositories > Indian Pacing and Electrophysiology Journal
Equipment and Supplies > Electrodes > Defibrillators > Defibrillators, Implantable
Pathological Conditions, Signs and Symptoms > Pathologic Processes > Arrhythmia
ID Code:2683
Deposited By:Indian Pacing and Electrophysiology Journal
Deposited On:07 March 2008

Archive Staff Only: edit this record